|
POWERING
UP AND FIRST-TIME
ELECTRICAL TESTS
When the Sentinel is first assembled, there are a few
tests which should be performed to ensure that all systems
are functioning properly. These tests are listed below,
in the order they should be performed.
1. POWER ON
Turn the Sentinel's main power on using the switch
located on the right side of the Starlite Colorhead.
This switch controls power to the entire system.
The first thing that should happen is a beep from the
keypad. The display will read "ZBE Inc." At
this time, the Starlite will proceed through a power-up
calibration cycle which takes about 10 seconds. This
ensures that the Starlite stays in perfect balance each
time it's turned on. During this time, press and hold
the key
on the keypad. This will cause the enlarger to skip
its initial referencing procedure temporarily while
the following tests are performed.
2. TEST MOTORS
This test, as are all the tests in this section is
accessed through the Sentinel's menu system. To get
to the Motors Test press the key. Select
"SENTINEL DIAGNOSTICS" from this Main menu
using the up and down ARROW keys under the numeric keypad,
and press .
The Diagnostics menu will now be displayed. From this
menu , select "TEST MOTORS" (again using the
ARROW keys) and press .
The display will now read "NEG: STOP LENS:STOP".
This is the Motors test screen. Use
, under
the number keys to jog the negative stage, and the
and in the lower right corner of the keypad to jog the lens
stage. When
or are pressed the appropriate stage should move in the direction
of the ARROW. This display for that stage should also
indicate the direction of movement: "UP" or
"DOWN". Both displays should read "STOP"
when no keys are being held down.
WARNING: It is important to note that the limit
switches are disabled in this test, so care must be
taken not to overrun the top and bottom limit switches,
nor to separate the bellows excessively. Damage to the
Enlarger may result.
Before ending this test, position the bellows so they
are separated by about 6 inches or so. This will aid
in some of the following tests. Press to exit
this test.
3. TEST ENCODERS
Access this test from the Sentinel diagnostics menu
in the same manner as the motors test was called up
in step 2, above. Select "Test Encoders" from
the diagnostics menu and press .
The display will now read "COL: nnnnn" on
the top line, and "BEL: nnnnn" on the bottom
line. (The "nnnnn" indicates a random number
displayed on the keypad.)
Gently pull straight up on the measuring cable
that is lying along the front of the column from the
negative stage up to the top of column. Do not pull
this cable more than 12 inches, or kink it while holding
it. As the cable is pulled straight up, both numbers
on the display should decrease in value. As the cable
is released, the numbers should count back up to their
original values, within a couple of counts. NOTE:
When releasing the cable, NEVER let it snap into position.
Next partially remove the bellows at the bottom. You
will see the bellows measuring cable inside at the front
left and rear right corners. Very carefully pull the
cable straight down (or as close to straight
down as possible) at the front left section of the cable,
do not pull down more than 8 inches. When doing this,
note the Lens Encoder number: it should decrease as
the cable is pulled, and return to its original value
when the cable is released.
If either number returns to a value grossly different
from where it started, or refuses to change at all,
there is a problem with the autofocus measurement encoders
system.
When finished with this test, press .
4. TEST LIMIT SWITCHES
Access through the Sentinel diagnostics as with the
previous tests. The display for this test is a readout
of the status of all three limit switches on the Sentinel.
The limit switches are used by the computer to tell
if any moving part of the enlarger is in danger of overrunning
its range.
Test the "Bellows" switch by reaching up
behind the bellows, under the negative stage, there
is a small lever located roughly at the centerline of
the bellows. Push this lever in, towards the rear of
the enlarger. When the switch clicks, the display for
the "BEL:" switch will go from off
to on as long as the plunger is held in.
Exit this test by pressing .
5. TEST REFERENCE
PULSES
The positional encoders in the Sentinel use reference
pulses to determine its initial position. Test the reference
pulses, using the "TEST REFERENCE PULSES"
in the diagnostics menu . The display should read: "COL:
---"on the top line, and "BEL:---" the
bottom line.
As in the encoders test, pull each cable slowly. In
this test, when a cable is pulled, a beep will occur
approximately every 4 inches of travel. At each beep,
a number should appear next to the appropriate encoder's
display. The cable along the front of the column is
the COL encoder, while the cable inside the bellows
is the BEL encoder.
When both encoders are made to beep and give a number
display, press to exit
this test.
This completes the initial testing of the Sentinel.
The main power should now be turned off, then
back on. This will put the Sentinel into its
normal daily power up procedure.
The colorhead will again cycle through its power-up
calibration. Then the display will read "PRESS
ANY KEY TO BEGIN REFERENCING." At this time make
sure there is nothing on the baseboard which will obstruct
the stages as they move to the bottom of the column.
Check for cords lying right next to the column. When
all is clear, press a key. The stages will move together,
then both will move to the bottom of the column. The
negative stage will then rise slowly until a reference
pulse is read from each encoder. When this occurs, the
keypad will display its main screen, showing the colorhead
filtration settings, exposure time, etc.
The Sentinel Enlarger is now fully installed and operational,
ready for the lenses and other items to be setup and
calibrated. Refer to the Getting Started section in
this manual to set up the Sentinel for everyday use.
THE
SENTINEL CONTROL SYSTEM
The Sentinel is a fully-automated photographic enlarging
system. As such it is computer-controlled. This section
offers a simple view of this computer control system.
In order to get the most out of your new enlarger it
is a good idea to understand the nature of the control
system.
The control system consists of two computers. One is
in the colorhead, the other is in the enlarger chassis.
The colorhead controller allows the Starlite to be one
of the most productive colorheads on the market, maintaining
constant filtration and density, exposure times and
print counts, and compensation for different paper types,
among many others.
The enlarger control system, or autofocus computer,
lets the machine focus and size your prints automatically,
even with different easels and lenses.
When these two computers are brought together, the
result is a powerful darkroom printing machine: the
Sentinel Enlarger.
The user need not worry, however, about learning to
use two computers! The Sentinel's two-part control system
is integrated together into one. All this is controlled
by an easy-to-learn keypad, which is fully backlit for
use in the dark.
This section outlines how the ZBE Sentinel control
system is operated by the darkroom operator.
USING
THE SENTINEL COMPUTER SYSTEM
The keypad which controls the enlarger features a number
of keys dedicated to operations
frequently used in routing printing. These include
keys for setting color filtration, print density and
exposure time, starting exposures, sizing a print, etc.
This makes most operations with the enlarger a simple
matter of a few keystrokes.
However, there are many more features built into the
Sentinel Enlarger. If each and every
feature had a button of its own, the keyboard would
become too large and complicated to use conveniently,
especially in the dark. To make things easier, these
features are accessed through a menu -driven software
system.
Please read the following to become familiar with how
the ZBE Sentinel system is operated.
GENERAL
USAGE OF DEDICATED KEYS
The operation of the Sentinel's keypad is simple. To
change any value accessed by one of the dedicated keys
( , , , , , , etc.),
simply press the appropriate key, enter the desired
numeric value, then press .
For example, say you want to change the exposure time
to 12.5 seconds:
- First, press
.
A cursor will blink next to the time displayed on
the keypad.
- Type the number of seconds, then tenths of seconds.
- For 12.5 seconds, type '1' , '2' , '5'. The decimal
point always stays at one place.
- When finished, press
. The new value is now entered in the computer.
All number entry in the computer system is done in
a similar manner. If a value has a decimal point in
it, the decimal will be fixed at whatever number of
places that particular value requires. Entering zeros
will move the number to the desired position if no fractional
value is needed.
If, during the entry of a number, you enter an incorrect
value and want to change it, entering a few zeros will
cause the number to blank back to zero, allowing a new
entry.
KEYPAD LAYOUT

GENERAL USAGE OF
MENUS
The menu system of the Sentinel Enlarger is organized
into sections, called sub-menu s. Each sub-menu , accessed
through the main menu , contains a group of items which
are related to one-another. For instance, all of the
functions needed to initially set up the Sentinel Enlarger
for autofocus operation are grouped in a sub-menu called
"SENTINEL CALIBRATION." This sub-menu appears
as one of the items in the Sentinel Main Menu.
Finding the item you want in the menu system is quite
simple. The keys involved are listed below and the function
of each is explained as it pertains to the menu system.
:
Puts the user in the Sentinel Main Menu from which all
other menus are accessed. When in a menu list, pressing
this key will advance the selection to the next item.
:
Select previous item in the current menu list. the
:
Select the next item in the current menu list. This
is the same as pressing while
in a menu .
:
All of the items in each menu are numbered. Pressing
a number key will move the current selection to the
item with that number. For experienced users, this is
a faster way to
get the desired item.
:
Activates the menu item currently selected. If that
item is a sub-menu (such as "SENTINEL CALIBRATION"),
that menu will now be displayed.
:
Exits the menu system and returns to the main screen
display.
The current selection is always displayed on the top
line of the display, with a '->' prompt next to it.
The next available selection is shown on the bottom
line. In the text of this manual, the word "select"
means to position the -> prompt next to a menu item.
The words "go into" mean to access that indicated
item by positioning the -> prompt next to it and
then pressing .
As an example of how this system works, suppose you
want to setup a new negative carrier for autofocus operation.
The function for this purpose is in the Sentinel Calibration
menu mentioned above.
- First, press
for the main menu . The Sentinel's Main menu will
now be on the screen.
- Use either
to move your selection to "SENTINEL CALIBRATION."
- Press
to go into that menu . The Calibration menu will now
be displayed.
- Move your selection to the item "SETUP CARRIER"
in this menu .
- Press
to activate that item. The computer will now run the
procedure for setting up a new negative carrier for
the autofocus system.
This was only an example, we don't want to actually
setup a carrier now. Press to return
to the main screen.
Understanding the menu system in the Sentinel Enlarger
will make using the machine easier and more productive.
It is worth the time to run through the menu s and familiarize
yourself with how they work and what they offer.
SENTINEL
ENLARGER MENUS LAYOUT

GETTING STARTED
BASIC OPERATION
This section covers topics necessary for the basic
operation of the Sentinel Enlarger. Such topics as focus
lamp control, entering color filtration and density
settings, and print sizing / cropping are covered here
from point of view for the first-time user. In this
section, we assume that the Sentinel is already mechanically
assembled and the first-time electrical tests have been
performed.
1. TURNING THE
POWER ON
The main power for the entire Sentinel Enlarger system
is controlled by a single switch, located on the right-hand
side of the Starlite 55 Colorhead.
When power is turned on the Sentinel's colorhead,
the Starlite 55, will go through a calibration procedure.
During this procedure, which takes about ten seconds,
the display on the keypad will display an opening message.
When the Starlite finishes, the Sentinel will take
over, the keypad reading "PRESS ANY KEY TO BEGIN
REFERENCING." At this time, press a key on the
keypad after checking that the baseboard area under
the lens stage is clear of any items which would collide
with the stage as it moves to the bottom of the column.
(grain focuser, etc.) Any cords or cables should be
kept away from the base of the column, as well.
During operation of the Sentinel, if access to the
inside of the Starlite is desired, the door may be opened
with main power on. Safety switches in the Starlite
will kill the power to the colorhead, leaving the Sentinel
power on. When the door is closed, the Starlite's power
will turn on again, and it will perform its ten-second
calibration procedure.
2. LENS SETUP
If the Sentinel is being used for the first time after
being assembled, the Autofocus/ Autosizing operation
will require that the lenses be set up. This is a one-time
procedure, so skip this section if your enlarger is
already set up. The Sentinel will not operate properly
unless each lens being used is set up according to the
following procedure. This initial lens setup procedure
entails focusing the enlarger at a small, then at a
large, magnification. From the magnification information
entered for each of these two sizes,the system will
be able to automatically focus at any print size along
the column. The procedure is as follows:
- At the main screen, rotate the lens turret to
select the lens to be setup, wait to hear the double
click from the keypad.
- Remove the easel if you are currently using one.
Select Easel, "NONE" in the Easel Selection
menu . (Press
,
then use
or
to select the "SELECT EASEL" menu option,
and press .
In this sub-menu , select "NONE" from
the list, and press
again.)
- Place the ZBE setup film in the 4 X 5 glass carrier
and install the carrier in the
enlarger. Select the 4 X 5 glass carrier from the Select
Carrier menu . (Press then use
or
to select the "SELECT CARRIER" menu option,
and press .
From this sub-menu , select the "4x5 G"
carrier from the list and press again.)
- Make sure that the Security Lock is off. This
item is controlled in the Sentinel Preferences menu
. (Press
, then use or
to select the "SENTINEL PREFERENCES" menu
option, and press . From this sub-menu , select "SET SECURITY LOCK",
and press again. If the display reads that the lock
is on, press to toggle it to the off setting. Press
when done.)
- Select "SETUP LENS" from the calibration
menu . (Press
,
then use
or
to select the "SENTINEL CALIBRATION" menu
option, and press . From this sub-menu , select "SETUP LENS",
and press again.)
- When prompted, enter the lens focal length. This
is used as name reference only; the true focal length
will be calculated automatically.
- Use the two sets of
, 's and to jog the enlarger into focus at a small size, which
is any position in the lower 10 inches of travel
of the enlarger. (The right-hand set of , 's are used to run both Negative and Lens Stages up or
down on the column for sizing.
or
under the Numeric keypad section move only the Negative
Stage for fine focusing.) When setting up large
lenses, do not jog to a size that is smaller than
magnification = 1.8, doing so will cause the focal
length to be calculated inaccurately.
- Use a grain focuser to achieve perfect focus,
moving the Negative stage with
or in the numeric keypad for fine focus. Single taps on
these keys are useful for very fine positioning.
Once on exact focus, press
.
- The display will now prompt you to enter the magnification
ratio. Measure the magnification using an accurate
ruler or scale. Measuring in millimeters makes calculation
of magnification easier. A transparent ruler works
best for easy reading. Measure the distance between
the appropriate alignment marks on the setup film.
Using the wrong marks will yield an incorrect measurement,
due to lens distortion. Suggested marks for various
lenses are:
o For 135mm to 150mm lenses use the 100mm marks.
o For 80mm lenses use the 50mm or 10mm marks.
o For 50mm lenses use the 10mm marks.
Accurate measurement of the magnification will make
or break the lens setup. Place the ruler on the
baseboard, aligning it to left the edge of the marks
projected onto the baseboard, then read the distance
to the left edge of the other mark. If your ruler
is marked in one millimeter increments, try to estimate
the distance to a tenth of a millimeter, that is
the accuracy necessary for a good setup.
Enter the resulting magnification ratio. The magnification
is the measured distance on the baseboard, divided
by the distance indicated on the setup negative.
For example, you may be setting up a 150mm lens
at about 1.9:1, you will be looking at the 100mm
marks projected down on your ruler. You may clearly
measure the distance as 190mm., but you must take
a guess down to a tenth, and you may see that the
measurement is closest to 190.5mm. In that case
the magnification you enter is (190.5mm/100mm).
After this magnification ratio is entered, the display
will read "FOCUS AT LARGE SIZE," then
press .
- Jog up the column to position within the top 10
inches of travel.
- Focus to grain sharp, as in step 7, and press
.
- Measure the magnification again and enter the
value. This is the same procedure as in step number
8.
After the large magnification is entered, the focal
length is calculated and the enlarger goes through a
Referencing procedure. After referencing, the lens setup
is complete and the lens will focus at all points along
the column. If the enlarger does not focus properly
or if the magnification displayed on the keypad is not
accurate, do the lens setup procedure again. Be particularly
careful about measuring and entering the correct magnification.
Even if the setup was done in perfect focus, a slight
error in measurement can cause it to be out of focus.
3. LAMP ON/OFF
CONTROL
The Starlite lamp may be turned on or off
using the LAMP +/- key. When the lamp is turned on,
the bulbs will light and the shutter will open simultaneously.
If the Starlite is in Filtered Light mode, the light
output will change to the color settings specified in
the upper left area of the keypad display. If the colorhead
is in White Light mode, the output will change to a
bright white light used for focusing or composing a
print.
To toggle between Filtered Light mode (for printing)
and White Light mode (for focusing), use .
If this key is pressed while in Filtration mode, the
White Light mode will be selected. At this time, the
lamp will turn on and the output will change
to white light. The lamp will now stay on until or is pressed. If
is pressed, the head will return to Filtration mode;
the light output will return to the specified color.
The lamp will then extinguish, the Starlite is now ready
for printing. If
is used, the lamp will simply turn off and the
head will remain in White Light mode with the upper
left area of the keypad display reading "FILTERS
OUT."
NOTE: In Filtered Light mode, there may be a
delay between the time when is pressed
and when the lamps actually go out. This is due to the
Starlite making fine adjustments to the color output.
This delay, which is usually a few seconds, is normal.
4. ENTERING
A COLOR FILTRATION
, , and
are used to enter color filtration values. Simply press the appropriate
color key, use
- to type in the desired color correction (CC) value, then
press .
When is pressed, the lamps will turn on and the head
will adjust to the new color, after which the lamp will
automatically go out. This will usually take a few seconds.
The Starlite is now ready for printing, that is, the
filters are in proper position.
As an example, suppose we want to enter a filtration
pack of Cyan 0, Magenta 30, and Yellow 30:
- From the main screen, press
.
A cursor will appear next to the 'C' in the upper
left of the display.
- Type the desired numeric value for Cyan, in this
case,
. At this point, you may either press
or another color key
( or
) to lock in the new value for Cyan. In this case,
press .
- The cursor will now move next to the 'M' to the
right of the Cyan display. (The Starlite's lamp will
come on briefly as the Cyan value is locked
in by the computer.)
- Enter a numeric value for Magenta. In this case
it is 30, so press
then
. When finished, press ,
since we want to enter a value for that color as well.
- The lamps will come on again as the Magenta
setting is locked in. The cursor will now move next
to the 'Y' to the right of the Magenta display.
- Enter a numeric value for Yellow, in this case 30.
Press
and .
- Since this completes our color entry, press
. The lamps will come on again as the Yellow setting
is locked in. When the lamps extinguish, the colorhead
filtration is ready for printing.
5. ENTERING DENSITY
The Starlite 55 Colorhead features Neutral Density
control which allows it to control the intensity of
its light output internally. This allows the Sentinel
Enlarger to make prints of different sizes without probing
the light at the baseboard or having to change lens
aperture or exposure time. This is normally done using
lens aperture, but the Starlite can do this accurately
and automatically. The Starlite's density value is printed
out in the main display below the color filtration.
The density value, like the color filtration values,
is in standard color correction units (CC's), with 30
points = one f-stop. The entering of a density value
is similar to color entry: press , type
the numeric value, then press .
or in
the numeric pad may also be used to increase or decrease
the density by one CC. When the density value is entered,
the colorhead's lamps will turn on and the intensity
of the output will be changed to new density. If the
Starlite cannot change to the new density value, the
exposure time will be changed to compensate. Such a
case would involve a density value which is either too
bright or too dim for the Starlite to output accurately.
6. SIZING
AND FOCUSING CONTROL
The Sentinel is a full-featured autofocus and autosizing
enlarger. For proper operation of this system, the lens
being used must be set up. (See Part B: LENS SETUP.)
In Autofocus mode, the enlarger will stay in focus at
all points up and down the column. The stages may be
moved either by entering a print size or magnification,
or by using the right-hand set of
or to jog the machine UP or DOWN. or
are useful for making fine adjustments when doing custom
cropping of a particular print.
There are two modes of operation for the Sentinel,
the Autofocus mode and Manual-focus mode. To change
from one to the other, use . The
upper-right area of the keypad display will read "MANUAL"
in the Manual-focus mode, or "SIZE:" in the
Autofocus mode.
In Autofocus mode, the right-hand
, , are used for jogging the enlarger onto the desired printsize
as they are in Manual-focus mode. may also be used in Autofocus mode. or
below the numeric keypad are inoperative in Autofocus
mode as focusing is accomplished by the Sentinel Enlarger's
control system.
In the Manual-focus mode, sizing and focusing the Sentinel
is the same as in the lens setup procedure. The right-hand
, 's are the main zooming and sizing control keys, while ,
below the numeric keys are used for focusing. is used somewhat like a Shift key on a typewriter to increase
the speed at which the enlarger moves and may be used
with either set of
, 's.
The other option for sizing in Autofocus mode is to
directly enter the desired print size or magnification
ratio. To do this, press . The
keypad will display a prompt for the entry of "SIZE:".
Press to toggle
between magnification or print size entry. When entering
a print size or a unit of measure, "in" for
inches or "mm" for millimeters, will be displayed
to the right of the number entry and may be toggled
by pressing
. In either mode, enter the numeric value desired, then
press . The printsize displayed here is calculated from the negative size
currently entered in the system; the units of measurement
are independent of the neg size. Whether jogging with
or
or directly entering a print size, the "SIZE:" display
will be updated as the enlarger moves.
For purposes of computing a print size, or for the
machine to jog to the correct size, the system must
be told the size of the negative (or distance between
cropping marks). This is a simple menu entry. To do
this, press
, select "NEGATIVE SIZE" from the main menu
, then press . A prompt will the appear for entering a new negative
size. This number will be multiplied by the current
magnification ratio to determine the print size. This
entry may be done in either Inches or Millimeters using
at this prompt.
If you have the size to density preference set to active
in the Starlite Preferences menu you will notice that
the light comes on and changes a few seconds after the
enlarger stops at a new position. This is the size to
density correction in the Starlite adjusting the lamps'
density up or down to achieve a constant print density
regardless of what size the enlarger is at. This automatic
correction only happens when the Size-to-Density Correction
is set to "active" in the Starlite Preferences
menu. The Size-to-Density Correction feature acts like an "imaginary user"
taking a Density Probe reading after each size change.
This feature makes it unnecessary to probe or re-test
for changes in print size. Note that if, for some reason
you want to use the D-Probe immediately after a size
change (for flesh-tone matching, etc.) you must let
the Size-to-Density correction take effect first (it
takes a few seconds) before taking any probe readings.
(Taking a probe reading too soon will cause both the
Size-to-Density correction and the D-Probe to compensate
for the same size change, causing a large over-correction.
If this happens, simply re-probe to correct the situation)..
7. KEYPAD
BACK LIGHT CONTROL
The Sentinel's keypad is fully backlit for use in the
dark. However, during the handling of photographic paper
and while printing, the light from the keypad may not
be desired. To extinguish the keypad's lighting, just
press
. This will toggle the backlights off and on.
In addition to the on/off control, the intensity
of the backlighting may be adjusted. A small knob is
provided under the right end of the keypad unit for
this purpose.
8. CARRIER LOADING
The negative carriers for the Sentinel slide into the
slot just below the Starlite. The carrier may be inserted
into the front or the left side of the enlarger. All
carriers supplied with the Sentinel are hinged in the
back and feature magnetic closure for non-slip negative
holding.
At times during printing, the carrier may difficult
to access due to the height of the enlarger's negative
stage, especially when doing large prints. To alleviate
this problem, the Sentinel features a programmable Load
Position. To program this Load Position, move the enlarger
to the desired location where carrier loading is to
take place. Then, in the main menu , select "CALIBRATION"
and press
. From the "CALIBRA-TION" sub-menu , select
"SET LOAD POSITION" and press .
The Load Position is now set and may be recalled at
any time during normal operation by pressing and simultaneously. The enlarger will move to the programmed carrier
Loading Position. When loading of the carrier is finished,
will
return the enlarger to its previous position.
9. VIEWING
THE AUTOFOCUS SETUP
Press to display
the current Lens, Carrier, Easel and Negative size currently
in use. Use this feature for double-checking which Carrier
and Easel are selected. This can be useful if the enlarger
is not focusing correctly because the correct Carrier
or Easel is not selected.
10.SAVING A
JOB TO MEMORY
The Sentinel has 200 job memory channels that can be
used to store specific print jobs. The information stored
for each job is the following: D,C,M, and Y, exposure
time, Emulsion number, f-stop, print size/magnification,
negative size, lens, easel, and carrier. The f-stop
is used only for the operator's reference and is not
used by the computer in any calculations. To save a
job, press . You
will be prompted for a job number; enter a five digit
number of your choice. If that number is already in
use you will be asked if you would like to over-write
the old job information. Next, you will be prompted
for an f-stop value. You may enter a f-stop value if
you wish, it will be displayed when the job is recalled.
While doing test prints, it is a good habit to save
the job after each exposure to ensure that the latest
data is always saved.
11. RECALLING
A JOB FROM MEMORY
To recall or load a job that was previously saved press
. You will be prompted to enter a job number. If the computer
cannot find a job by the number entered, the display
will show "DOESN'T EXIST, TRY AGAIN." When
the job is recalled the light will turn on and the filtration
will be put in and the recalled f-stop is displayed
for a few seconds. If the correct lens is not currently
selected, you will be prompted to move the correct lens
into position. The enlarger will move to the new printsize.
Then the display will go back to the main screen and
the light will turn off. The enlarger is now ready for
printing.
12. MAKING
A TEST EXPOSURE
Making a test exposure with the Sentinel is similar
to other enlargers. The Sentinel has the capability
to remember print setups, so that once a test print
is successful, it need not be done again. A procedure
for making a test print is outlined here:
- Insert the negative into the carrier and install
the carrier into the enlarger.
- With the colorhead in Filtration mode, turn the
Colorhead lamp on by pressing
to get White light. Use
or
to crop the print to the desired composition. When
finished, turn the colorhead lamp off by
pressing
again. This will return the colorhead to the Filtration
mode.
- Enter the desired first-time print data for Cyan
(usually zero), Magenta, Yellow, and Density.
- Enter the desired exposure time.
- Just before making the actual print, save the
setup under a job number in the Sentinel's memory.
This way the setup will be saved if the machine
is used while this test print is in process (perhaps
for another test exposure).
- Place the paper in position and make the print
by pressing
.
This is one round of the test printing process. Several
tests may be required to achieve a perfect print. Once
the perfect print is made, this final setup should be
saved so that it may be recalled from memory if reprints
of the negative are required in the future.
SENTINEL
PREFERENCES
1: SET LOAD POSITION
Memorizes position for carrier loading.
2: SET TOP LIMIT
Memorizes enlarger top limit if lower than factory
setting.
3: CLEAR TOP LIMIT
Clears limit set above.
4: SET BELLOWS LIMIT
Memorizes maximum bellows extension limit.
5: CLEAR BELLOWS
LIMIT
Clears limit set above.
6: ENTER ENLARGER
ID NUMBER
Sets number used to identify enlarger on lab network.
7: SET SECURITY LOCK
Toggles security lock mode for doing Sentinel calibrations
and preferences.
The Sentinel and Starlite both have a set of options
called "Preferences" which can be set to meet
the specific needs of your lab. These preferences are
accessed in the Sentinel Preferences and Starlite Preferences
menus. The Sentinel preferences relate to the enlarger
movement up and down the column and the Starlite preferences
relate to the colorhead operation. The Sentinel preferences
you should consider during initial setup are the following:
1: SET LOAD POSITION
2: SET TOP LIMIT
7: SET SECURITY LOCK
Before these preferences can be setup, you must turn
off the security lock. The Set Load Position preference
is used to memorize the position you would like the
enlarger to move to for ease of carrier loading. To
set the position, jog the enlarger up or down the column
using or
. Then select the Set Load Position option. Now when you press and simultaneously, the enlarger will move to that position while you
load or unload the carrier. When finished with the carrier,
press and
the enlarger will move back to previous position, ready
for printing.
The Set Top Limit preference is used if the enlarger
is setup in the darkroom such that it can hit the ceiling
when run to the top of the column. To prevent this,
jog the enlarger to a position at least three inches
below the ceiling, then select the Set Top Limit preference
to memorize the new top limit. For information on other
Sentinel preferences, please refer to the Sentinel preferences
chapter in the Reference Section of this manual.
The Starlite preferences you should consider during
initial setup are the following:
1: SIZE TO DENSITY CORRECTION ACTIVE
2: DENSITY SLOPE DIRECTION Z POSITIVE
3. TIME TO DENSITY CORRECTION ACTIVE
These preferences must be set properly before you align
the Starlite to a video analyzer or else the Starlite
will not function properly with the analyzer. To change
a preference, press , when the appropriate preference is displayed, press
the number keys to change the preference and then press
to
lock in the new preference. Press exit
the menu .
The Size to Density Correction preference applies to
the Starlite receiving print size information from the
Sentinel and adjusting the light output to maintain
constant print density as the print size changes. When
Active, this preference will cause the Starlite to adjust
the light intensity automatically when the enlarger
moves to a new position. Approximately 4-6 seconds after
the enlarger stops moving, the lamps will come on,
dim or brighten and then go off. NOTE: the
adjustment will not be made if the colorhead is in Filters
Out mode or the enlarger is in Manual-focus mode. This
preference should always be active unless you specifically
want to make size density changes manually by making
test prints.
The Shutter preference disables or enables the internal
rotary shutter. This should always be set "ACTIVE"
unless there is a problem with the shutter.
The Density Slope Direction is a very critical preference
selection. This determines whether the density values
used in the Starlite increase or decrease with increased
light output. For example, if your video analyzer increases
the density number for a darker print, say from 64 to
70 density points, then you should set the density slope
direction to positive so an increasing density value
increases light output and thus increases print density/darkness.
Most video analyzers use positive slope. If you are
not using a video analyzer choose the mode that you
will be most comfortable working with.
The Time to Density Correction preference ties the
exposure time to the density value so when the time
is increased the density value is increased. Also, if
a density value is entered that is either too bright
or too dim for the colorhead to achieve the time will
be adjusted up or down to achieve the desired density.
This mode should be set to Active, in most cases, especially
when a video analyzer is used.
The Maximum Lamps Power preference sets the maximum
amount of power that will be applied to lamps. The range
is from 60 to 99%. If set to 99%, the lamps will be
allowed to run at the maximum possible power level.
However, a power level of a few percent less will greatly
increase lamp life. The factory setting is 90%. If you
want the absolute greatest amount of light output, you
should set this preference to 99%. For information on
other Starlite preferences, please refer to the Starlite
preferences chapter in the Reference Section of this
manual.
CARRIER SELECTION
To select a carrier offset use Sentinel Main menu option,
"SELECT CARRIER." Before using a new carrier,
the carrier must be setup using the Carrier Setup procedure
(See next section). Select the appropriate carrier from
the list shown in this menu function. The carriers are
listed by format size for convenient reference. Use
or
to move up and down the list, or press - to jump
to the corresponding carrier in the list. Press
when the correct carrier entry is selected.
CARRIER SETUP
Each carrier must be setup for proper Autofocus operation.
The one exception is the 4 X 5 glass carrier, we recommend
you do not change the setup for this carrier and keep
it as the Zero Offset carrier that is always used for
lens setups. Each carrier may have a slightly different
thickness from it's bottom surface to the emulsion surface.
A difference of as little as a thousandth of an inch
can make a difference in grain sharpness. To correct
for these variations, setup the carriers using the following
procedure:
1. Select the shortest lens available (the 50mm lens
you've already setup).
2. Install and select the carrier you wish to setup.
Select easel "NONE" using the Select Easel
item in the main menu and remove any easel from the
baseboard.
At the main screen jog the enlarger to 3 inches below
it's maximum height, in Autofocus mode.
5. Select Setup Carrier from the Calibration menu
.
6. Select the appropriate carrier from the displayed
carrier menu .
7. Tap
- and use a grain focuser to achieve a grain sharp image.
8. Press
when grain sharp.
The Carrier Setup procedure is now complete.
EASEL SELECTION
When changing the easel being used with the Sentinel
Enlarger, that change must be entered in the Sentinel's
computer. This is simply a matter of selecting the current
easel from the list using the Select Easel item in the
main menu . When in this menu item, you will be presented
with a list of available easels, starting with "NONE"
and then letters
'A' through 'H'. Selecting "NONE" will load
an easel offset of zero which means no easel is being
used.
For accurate autofocus and autosizing each easel must
be designated with a letter and setup using the Easel
Setup procedure (See Sentinel Calibration menu section).
EASEL SETUP
Before a new easel is used with the Sentinel Enlarger,
it must be setup to allow the autofocus system to properly
calculate the image position when the easel is in place.
To program an easel follow these steps:
- Select the shortest lens available (the 50mm lens
you already setup).
- Install and select the 4x5 glass carrier with
the ZBE setup film.
- Select easel "NONE" using the Select
Easel item in the main menu.
- At the main screen jog the enlarger to 3 inches
below it's maximum height, in Autofocus mode.
- Remove any easel from the baseboard.
- Place a large piece of white paper on the baseboard
and mark where the 10mm alignment marks from the
Autofocus setup negative are projected onto the
paper. Do this carefully, using very fine lines.
- Now select "SETUP EASEL" from the Calibration
menu .
- You will be presented with a list of easel "NAMES"
to choose from. Select 'A' if it is the first easel,
and press
. Write the letter 'A' on your easel for future reference.
- Place the easel on the baseboard.
- Place the paper with the marks on the easel.
- Now use
and
to jog the enlarger up or down until the 10mm marks
exactly line up with marks you made on the white
paper.
- Press
when the marks are perfectly aligned.
The Easel Setup procedure is now complete. This easel
may now be selected from the Select Easel menu item,
in the Sentinel main menu , when its use is desired.
KEYPAD
LAYOUT

|