Technical Support Document 036 - 5/97
THE MANUAL CONTROL TEST
This test allows the user to run various systems in
the Starlite individually, and without computer intervention.
It will test Filter Blade and Shutter movement, Color
Sensor response, as well as Lamp Voltage Regulation
and Control.
To enter this test, go into the Diagnostics menu, and
select the first item, MANUAL CONTROL TEST. (Press MENU,
<8>, ENTER, ENTER.) Key responses in this test
are very slow, so press any keys used for testing very
slowly.
While in the test, the screen will appear as shown
below. The various portions are explained below, followed
by the actual test routines.

1 Base Filtration Values. These are the Cyan,
Magenta, and Yellow filtration values as read directly
from the color sensor. These are raw values, and have
no offsets added to them.
2 Display Filtration Values. These are the D,C,M,Y
values as shown on the main screen during normal operation.
They are calculated using all offsets in effect, including
the Master Color Balance Offsets, Emulsion Offsets,
etc.
3 Color Sensor Voltages. These are the voltages,
in millivolts, that the color sensor is actually putting
out. The Starlite uses a logarithmic light sensor, so
these voltages will not
reach zero when the lamps are turned off.
4 Lamp RMS Voltage. This is the amount of AC
mains voltage being sent to the lamps. This voltage
will range between about 7 volts up to the rated voltage
of the lamps in the Starlite.
I. FILTER BLADE MOVEMENT
In the test screen, the number-pad keys are used to
actually move the filters. The keys are operated like
up/down arrows, not as number entry keys. The chart
below shows the usage of keys to move the filters and
shutter.
| CYAN Filter IN |
1 |
2 |
3 |
CYAN Filter OUT |
| MAGENTA Filter IN |
4 |
5 |
6 |
MAGENTA Filter OUT |
| YELLOW Filter IN |
7 |
8 |
9 |
YELLOW Filter OUT |
| SHUTTER Close |
^ |
0 |
v |
SHUTTER Open |
For Example, press and hold the <3>
to pull the Cyan filter all the way out of the light
path.
To check filter movement:
- Open the Starlite's door and tape the safety interlock
switches so that the head may be run with the door
open. Leave the mixing chamber in place while the
head goes through it's power-up procedure.
- Turn the power on in the normal manner.
- When the Colorhead reaches the normal operations
screen, access the Diagnostics Menu for the MANUAL
CONTROL TEST, as described in the beginning of this
document.
- Once in the test screen, remove the mixing chamber
to give a better view of the filter blades.
- Move the Yellow filter blade all the way out of
the light path by pressing and holding the <7>
key. While doing so, look at the blade itself to see
if the movement stops before reaching full open. Also
listen for any excessive noise while the filter is
moving. The movement should be smooth and even, with
no binding or hesitation.
- Repeat step #5 for the Magenta then the Cyan Blade.
Use the <6> and <3> keys, respectively.
- Next, try the shutter. Pay extra attention to the
shutter's movement, since most failures associated
with blade movement are associated with the shutter.
( It does the most work in normal operation.)
- If the movement of all three blades and the shutter,
out of the light path, is clear, proceed to close
all the filters, in turn.
- Start by closing the shutter, then proceed with
Cyan, Magenta, then Yellow. This will allow a clear
view of each blade as it closes.
This completes the visual test of the filter blade
movement. If everything checks out, please proceed to
the next section.
II. COLOR SENSOR RESPONSE
The test for the color sensor is essentially the same
as that for the filter movement. However, it is done
with the mixing box in place. Since this test relies
on the fact that the filter movement is correct, the
previous test should always be done prior to this one.
The theory of this test is that each filter will change
the amount of light for it's particular color by about
220cc's, from fully-opened to fully-closed. The color
sensor should be able to read this range of light output
for each channel. Sensor stability will also be examined.
During this test, please refer to the screen layout
and number-pad function chart at the beginning of this
document, if needed.
To test sensor response:.
- Install the mixing chamber into the Starlite enclosure
in the normal manner.
- Check that there is tape over the safety interlock
switches, and turn the power on.
- After the Starlite has reached the main operating
screen, access the Diagnostics menu and activate the
MANUAL CONTROL TEST function.
- In this test screen, move all filters all the way
out of the light path. The shutter must also be open.
- Turn the Starlite's lamps on by pressing the LAMP
key. Key response while in this test mode can be sluggish,
so press keys slowly.
- The upper, left set of numbers on the screen will
start to indicate the filtration of light being read
by the color sensor.
- Once these numbers stabilize, take a note of each
of the three values. The actual values are not important,
but all should usually be below 100.
- Watch the numbers for signs of excessive movement.
> The numbers should maintain their values within
about 2 or 3 points.
> If there is a stability problem with the Starlite's
color sensor, the numbers will fluctuate by 5 points
or more.
> Fluctuations less than 5 points, yet more than
two, usually indicate a lamp voltage stability problem.
(This will be tested in the next section.)
- If the numbers appear stable, continue the test
by closing each filter, in turn, with the lamp still
on. As each filter moves, it's corresponding number
should climb to a positive value that is at least
190 cc's more than the value noted in step #7. They
are usually 220-230 point different. Also check that
each filter affects only it's particular color by
a large amount.
- After all three filters are all the way in the light
path, watch them once again for stability. They should
not fluctuate any more than they did in step #8.
If all of these things check out all right, the Starlite's
color sensor is functioning properly.
III. LAMP VOLTAGE REGULATION AND CONTROL
The Starlite's lamp voltage regulation and control
is controlled by the computer. The maximum and minimum
lamp voltages are set when the Starlite is turned on.
The computer allows the user to monitor and test the
performance of the lamp control circuitry from the Manual
Control Test screen.
This test is a simple matter of running the lamps dimmer
and brighter, comparing the light output with the voltage
display.
The voltage going to the lamps during the Manual Control
Test is shown in the lower, right-hand corner of the
keypad display. The voltage is controlled using the
right-hand set of arrow keys on the keypad. These are
the set of arrow keys directly under the Enter and Cancel
keys.
To check lamp voltage control.
- In the Manual Control Test screen, turn the lamps
on by pressing the LAMP key.
- The indicated voltage on the lamps should read about
73 volts on 110-volt units, or 107 volts on 220-volt
units. (These voltages are for a "Maximum Lamp
Power" setting of 90%. Please see the Starlite
user's manual for more information on this "Preferences"
setting.)
- Watch the lamp voltage number for any signs of instability.
This number should not fluctuate more than +/-1 volt.
- Press and hold the down-arrow under the Cancel key
on the keypad. The lamps should visibly get dimmer,
in steps, and the indicated voltage should decrease
as well. Pay attention to this behavior. the lamps
brightness should decrease, while the voltage readout
also goes down. Any behavior to the contrary would
indicate a problem.
- After a while, the lamp voltage will reach it's
lowest value. For 110-volt heads, this will be about
7.5 volts. For 220-volt units, it will be about 11
volts. If it is unable to reach a value close to those
stated here, bottoming-out at a considerably higher
voltage, there may be a problem with either the system
board, or a failed power-on calibration routine that
set an incorrect low-voltage. This lowest voltage
should also match the voltage displayed when the lamps
are turned off.
If the voltage is easily controllable and stable, the
lamp circuitry is functioning properly.
At the completion of this last test, we have now checked
all of aspects of the Starlite's light control system
for both Density and Color.
|