Technical Support Document 041 - 5/97
REPLACING THE COLOR SENSOR IN THE ZBE STARLITE
55 COLORHEAD
The Starlite 55 Colorhead uses a Color Sensor to read
the filtration of output light directly from the inside
of the mixing chamber. The colorhead is thus able to
maintain consistent light output over a wide range of
filtration and density. Following are instructions for
replacing this sensor.
NOTE: THIS PROCEDURE REQUIRES A RECALIBRATION
OF THE STARLITE 55, AND WILL RENDER ANY CURRENT OR VCNA
COLOR STANDARD ALIGNMENT INVALID.
- Unplug the Colorhead from the Mains AC Power and
open the door. The color sensor faces the left-hand
wall of the mixing chamber. Remove the Mixing Chamber
at this time.
- Unplug the Sensor's Cable from the Main Circuit
Board at the rear of the interior of the Colorhead.
This is a flat grey 'ribbon' cable which plugs onto
the main board's "SENSOR#1" connector at
the lower left corner of that board. Note the red
stripe that runs along one side of the grey ribbon
cable. This lines up with pin one ("1")
of the board's connector. This pin is denoted by a
small white "1" printed on the green circuit
board itself.
- The sensor is attached to a metal mounting bracket
which is mounted to the left side of the colorhead
with four screws. Loosen these screws and remove the
Sensor from the head.
- If the new Sensor has no grey cable attached to
it, transfer the cable from the old sensor, making
sure to maintain the red-striped side of the ribbon
in the same orientation on the new sensor as it was
on the old one.
- Position the new Sensor in the Colorhead and attach
it using the screws removed in step.
- Plug the other end of the cable onto the Main Circuit
Board, with the red stripe closest to the end of the
connector with the "1" next to it. Be sure
the connector is properly positioned over both rows
of pins.
This completes the installation of the new color sensor.
It will now need to be re-calibrated for operation in
this particular Starlite. To do this, enter the "STARLITE
DISGNOSTICS" menu. From this menu select the "AUTOCALIBRATE
COLOR SENSOR" function, and press the Enter key.
The calibration will take about 15 seconds.
The Starlite 55 is now ready for normal use. Initial
test printing, Shirley alignment, etc., will now be
necessary as the Colorhead is now calibrated to a slightly
different color balance.
|