Technical Support Document 046 - 5/97
STARLITE 55 INITIAL ALIGNMENT, STEP BY STEP
PROCEDURE TO SET UP STARLITE WITHOUT A VCNA
Note: This Procedure assumes the customer has
difficulty setting up and will commence by Deleting
any figures deliberately stored or stored in error.
Commence by deleting any erroneous figures which may
be stored.
- Switch On the Starlite and wait for the self checking
procedure to terminate (approximately 10 Seconds).
- Press the MENU key
- Press the #7 key
- Press ENTER key
- Press the #7 key
- Observe the display and note the arrow points to
Master Reset
- Press ENTER key
- Observe the display indicating everything will be
erased
- Press ENTER key
- Observe display which now reads ZBE, Inc
- Switch Off Starlite using Mains On/Off switch. This
clears all memories of any erroneous data.
- Switch on Starlite
- Observe display which indicates the memory has been
cleared.
- Press ENTER key
- Observe display which indicates the Starlite has
a new EPROM
- Press ENTER Key
- The Starlite now calibrates the internal Sensor
which takes approximately 8 Seconds.
- Observe display which now shows standard default
format.
- Set the enlarger approximately 1/3rd up the column
and size and focus image of a 'Shirley Girl' reference
negative.
- Press MAGENTA key
- Press #2 and #0 keys
- Press ENTER key
- Notice the exposing lamp will switch On and color
will be inserted
- Press YELLOW key
- Press #5 & #0 keys
- Press ENTER key
- Observe exposing light is switched On and color
inserted.
- Press TIME key
- Press #5 & #0 Keys
- Press ENTER key
- Observe 5 Seconds has been entered as the exposure
time.
- Disregard any figure appearing beside the D: (Density)
- Stop down the enlarging lens to f11
- Switch Off the darkroom lights and make an exposure.
Process this test print on the most commonly used
photographic paper in the darkroom. This will later
be used as a reference for other paper emulsions.
- Continue making test prints until you are completely
satisfied the test print cannot be improved.
- Note: Figures used by Video analyzers can vary.
The figures used below are typical but may be different
in your case. Use the appropriate figures for your
lab. For the purposes of this Procedure we will use
the following figures:
Density = 64 / Cyan = 0 / Magenta = 30 / Yellow
= 30
- Press Menu key
- Press #6 key
- Press ENTER key
- Press #2 key
- Press ENTER key
- Press #6 & #4 keys
- Press LAMP key if the Density reading shows
a minus sign
- Press ENTER key
- Press ENTER key (This moves the cursor from
Cyan to Magenta)
- Press #3 & #0 keys
- Press ENTER key
- Press #3 & #0 keys
- Press ENTER key
- Now calibrate the Density Probe
- Press ENTER key
- Remove negative from carrier and replace carrier
- Press ENTER key
- After a few seconds the Exposing light will be extinguished
and the Calibration procedure is completed. Note:
Ideally this procedure will only need to be completed
once. Any future adjustments will be made elsewhere
according to your printing requirements.
|