Technical Support Document 055 - 7/97
CALIBRATING BLACK-AND-WHITE D-PROBE on the
Starlite 55 Colorhead
INTRODUCTION
When printing either color or black-and-white work,
it is generally desirable to align your lightsource
output to coincide with your lab's standard. Most labs
have some form of standard negative, which, in the day-to-day
working of the photo lab, will print at a specific Density
value at the enlarger.
The Starlite colorhead allows you to change its output
light intensity. In color mode, you are also able to
easily change the amount of light that goes with a particular
density value on the keypad, through the use of the
Calibrate Master Color Balance procedure. In Black-and-White
mode, this is a bit more tricky, and requires some trickery
to make the Starlite lightsource conform to your lab
standard.
THE IDEA
This idea takes advantage of the fact that the Starlite's
Density probe is designed to change the amount of light
coming out of the head that corresponds to the Density
value on the main screen. An example of the Density
Probe's purpose would be shown in the situation where
you just made a print of size 8x10, at a density setting
of 64 (for example), and wish to make the same print
at 5x7. Once the 8x10 is made, you move the enlarger
and focus at the 5x7 size, but the settings at the lightsource
would make a print that is too dark. So you use the
density probe to show the Starlite the light at the
baseboard. The Probe will see the brighter light, and
once it is activated, will allow the Starlite to dim
the light accurately in order to make a good print at
the new size. When this is done, the Starlite is now
putting out less light at the same density
value!
The above example works the same for both black-and-white
and color printing modes in the Starlite. However, the
setting of that value of 64 is only directly available
in color mode. In black-and-white, we will need to use
the Density Probe's ability to change the amount of
light that goes with the keypad density number, to make
the density number agree with your photo lab's standard.
THE PROCEDURE
The alignment procedure that follows depends on a few
settings in the Starlite Preferences Menu to be at particular
values. The settings that are important are:
Name of setting in menu: Setting Value:
4: TIME TO DENSITY CORRECTION ACTIVE (This is the default
from factory)
9: PROBE WITH AUTOCORRECTION ACTIVE (This is the default
from factory)
10: LAMP POWER DISPLAY ACTIVE
For information on how to change these settings, please
refer to the Starlite User's Manual. Of course, the
Starlite should also be in Black-and-White Printing
Mode.
To align the Starlite to your lab standard you
will need the following:
Your normal reference ("Shirley") negative
for black-and-white. The Density Value at which you
want the Starlite to make a perfect print from this
negative.
The procedure is as follows:
- Compose and focus your enlarger to make the standard
print. (An 8 x 10 print works well in most cases.)
- Remove the negative from the carrier.
- Select a suitable grade for the standard negative.
- Enter the Density Value for the standard negative
into the Starlite keypad. Chances are that the Starlite
will beep, and the exposure time will change.
If the exposure time decreases, you
will need more light from the Starlite as a
result of this procedure. If the exposure time
increases, you need less light than
you are getting for your standard Density Value. This
is the most common case.
If you need less light, close your lens almost all
the way. If you need more light, open it almost all
the way.
- Re-enter the exposure time to the value at which
you want to do your printing. This will cause the
density value to change, but ignore it.
- Temporarily calibrate the probe at this light level.
To do this, press MENU, then <6> for "STARLITE
CALIBRATION," and press ENTER. From Starlite
Calibration, press <6> for "CALIBRATE B/W
PROBE DENSITY," and press ENTER again. The light
will come on and the screen will tell you to remove
the negative and take a probe reading. Place the probe
directly under the lens, and press it's button, making
sure your hand is not blocking the D-Probe's "eye."
The Density Probe is now calibrated to this new light
level.
- If you had opened the lens in Step #6, close it
now by about 2 full f-stops. If you had closed the
lens in Step #6, open it by about 2 full f-stops.
- With the new lens setting, take a D-Probe reading
by placing the Probe directly under the lens, and
pressing the button. The lamps will come on, then
change intensity, then turn off.
- Once the probing is finished, again enter your desired
target density value.
If the exposure time changes, repeat this procedure,
starting at Step #5.
If the exposure time does not change, the head is
now capable of being set to your standard Density
Value. The next thing to do is to make a good print
at this density value.
- At the current settings, place a piece of photo
paper, and make a print in the normal manner. Most
likely, this print will be off in density, and maybe
even in contrast grade. Correct the grade by changing
the keypad Grade Value. Correct the Density of the
print with only the lens aperture. DO NOT CHANGE
THE DENSITY VALUE OR EXPOSURE TIME ON THE KEYPAD AT
THIS TIME. Do not worry: after this procedure
is complete, you will be able to print at whatever
lens settings you wish.
- Once the print looks good, we can re-calibrate the
Density Probe once more to lock in the final setting.
Calibrate the Probe as done in Step #8. This will
let the Probe know that the current light level is
what the Starlite should have on the baseboard at
the current Density setting.
After doing this procedure, you can print at whatever
lens aperture you wish, all you have to do is to make
sure you Probe after changing it!
If you have questions or problems with this procedure,
please contact ZBE at the email, address or phone numbers
below.
|